February 2020
PRODUCT NAME | LIST NUMBER |
---|---|
i-STAT CHEM8+ cartridges (blue) | 09P31-25 |
i-STAT CG4+ cartridges (blue) | 03P85-50 |
i-STAT G3+ cartridges (blue) | 03P78-50 |
Dear Valued Abbott Point of Care Customer,
This letter contains important information regarding the use of capillary whole blood to run tests with the three blue i-STAT® cartridges listed above.
Please review the information contained in this Product Correction letter, as our records indicate that your facility utilizes one or more of the i-STAT cartridges associated with this communication.
If you are using the white i-STAT CG4+ cartridges (list number 03P85-25) or the white i-STAT G3+ cartridge (list number 03P78-25), this notification does not apply to those cartridges.
Abbott Point of Care is communicating this information to all potentially impacted customers.
The three i-STAT cartridges listed above may currently be used for the quantitative measurement of sodium, potassium, chloride, urea, ionized calcium, TCO2, glucose, creatinine, lactate, pH, pCO2, pO2 and haematocrit in venous, arterial or capillary whole blood.
It has been determined that Abbott Point of Care will no longer support the use of capillary whole blood samples with the i-STAT blue CHEM8+, blue CG4+ cartridges, and blue G3+ cartridges listed in the table above. Therefore, Abbott Point of Care has decided to remove capillary whole blood as an acceptable sample type from product labelling for the three cartridges listed in the table above.
This action is not being done to reduce risk to health. Abbott Point of Care has not received any reports of patient harm associated with the use of the three cartridges listed above.
If you are running exclusively venous and/or arterial samples on the above i-STAT cartridges, your facility is not impacted; however, the use of capillary samples with the above i-STAT blue cartridges will no longer be an acceptable sample type.
If capillary whole blood is a required sample type at your facility for the cartridges listed above, it is recommended that your facility evaluates the potential to transition testing to an alternate i-STAT cartridge or an alternate test method that supports the use of capillary whole blood. Abbott will notify customers when the products are available for use with capillary whole blood again.
The listed cartridges may continue to be used with venous and/or arterial whole blood samples.
Please note the white i-STAT CG4+ cartridges (list number 03P85-25) and the white i-STAT G3+ cartridge (list number 03P78-25) are not affected by this Product Correction Notification, and capillary samples will continue to be an acceptable sample type when used with these white cartridges.
If test results obtained with an i-STAT blue CHEM8+, blue CG4+, or blue G3+ cartridge appear inconsistent with a patient’s clinical assessment, the patient sample should be retested using another cartridge or an alternate test method. Please contact Abbott Point of Care Technical Support for support and to report the information.
If you have forwarded any of the blue i-STAT cartridges listed in the table to another facility, we request that you please provide a copy of this letter to them.
Product labeling for the three blue cartridges listed in the table will be updated to reflect the removal of the capillary claim. The updated labeling will be posted on the Abbott Point of Care website (www.pointofcare.abbott).
If you have any questions regarding this information, please contact your Abbott Point of Care support services representative.
Abbott Point of Care sincerely apologizes for any inconvenience this may create for your facility. We appreciate your understanding in this matter and are fully committed to supporting you moving forward.
©2025 Abbott. Med enerett. Med mindre annet er angitt, er alle produkt- og tjenestenavn som vises på dette nettstedet, varemerker som eies av eller lisensieres til Abbott, dets datterselskaper eller partnere. Ingen av Abbott sine varemerker, handelsnavn eller kjennetegn på dette nettstedet kan brukes av andre uten forhåndsinnhentet skriftlig tillatelse fra Abbott, bortsett fra der de brukes for å identifisere produktet eller tjenesten til selskapet.
Dette nettstedet er regulert av gjeldende lover og statlige forskrifter i USA. Det kan hende at produktene og informasjonen som finnes her, ikke er tilgjengelig i alle land, og Abbott påtar seg ikke ansvar for informasjon som muligens ikke er i samsvar med juridiske prosesser, reguleringer, registrering og bruk i det enkelte land.
Din bruk av dette nettstedet og informasjonen som finnes her, er underlagt nettstedets bruksvilkår og retningslinjer for personvern. Bildene som vises, er kun til illustrasjonsformål. Personer på bildene representerer ikke virkelige personer. Erklæring om personvernforordningen (GDPR).
Alle produkter er ikke tilgjengelige i alle regioner. For informasjon om tilgjengelighet i bestemte markeder kontakter du din lokale representant. Kun til bruk ved in vitro-diagnostikk. Ønsker du informasjon om i-STAT-testkassetter og tiltenkt bruk, kan du se de enkelte produktsidene eller informasjonen om kassetten (CTI/IFU) i delen for i-STAT-support.
Abbott – ledende innen pasientnær diagnostikk