Fast and precise determinations of activated clotting time (ACT) are essential during cardiopulmonary bypass (CPB) surgery to monitor the degree of heparin-induced anti-coagulation. An inaccurate ACT puts patients at risk for thrombosis and hemorrhage and risks damage to the extracorporeal circuit.
This white paper describes the rationale behind the implementation of the i-STAT System at Augsburg Hospital, where mechanical ACT testing has been used for many years. Data, collected on site, are described which compare the precision of the i-STAT System to its traditional, mechanically based ACT testing device.



Rationale for i-STAT Implementation
Goal:
Improved efficiency
Improvement
Precision on the i-STAT gave clinicians confidence in the results. Repeat testing on comparator device added an additional 17 minutes in 7% of samples.
Goal:
Reduced QC workload
Improvement
- No testing interruption for QC testing
- i-STAT System QC easily customisable to be compliant with the RiliBAk system
- QC results sent directly to IT system creating a paperless documentation and verification
Goal:
IT connectivity
Compliance
Full integration into Augsburg Hospital's IT system.
Results of Data Comparison
- Of the duplicate samples TESTED WITHIN THE THERAPEUTIC RANGE (<450 s), 15% (8) had a duplicate difference of (>50 s) WITH THE comparator, compared TO 3% (1) of i-STAT duplicates
- In 7% of the duplicates on the comparator device, one of the duplicate measurements fell out of range (>1000 s) while the other one was in range. This did not occur on the i-STAT System
- The average difference between the i-STAT duplicates was 26.3 seconds, 30% better than the average difference of 37.4 seconds for the comparator device
View The Complete Case Study


Results of Data Comparison
- Of the duplicate samples TESTED WITHIN THE THERAPEUTIC RANGE (<450 s), 15% (8) had a duplicate difference of (>50 s) WITH THE comparator, compared TO 3% (1) of i-STAT duplicates
- In 7% of the duplicates on the comparator device, one of the duplicate measurements fell out of range (>1000 s) while the other one was in range. This did not occur on the i-STAT System
- The average difference between the i-STAT duplicates was 26.3 seconds, 30% better than the average difference of 37.4 seconds for the comparator device
View The Complete Case Study